Journal Home
Search for

Volume 50, Issue 4, Pages 730-737.e1 (October 2009)


View previous. 9 of 103 View next.

Intermediate results of a United States multicenter trial of fenestrated endograft repair for juxtarenal abdominal aortic aneurysms

Fenestrated InvestigatorsRoy K. Greenberg, MDaCorresponding Author Informationemail address, W. Charles Sternbergh III, MDb, Michel Makaroun, MDc, Takao Ohki, MD, PhDd, Timothy Chuter, MDe, Priya Bharadwaj, PhDf, Alan Saunders, MSf

Received 12 January 2009; accepted 19 May 2009.

Objective

This article reports the intermediate-term (24-month) outcomes of a prospective multicenter trial designed to evaluate the Zenith Fenestrated AAA Endovascular Graft (Cook Medical, Bloomington, Ind) for treating juxtarenal abdominal aortic aneurysms with short proximal necks. The study goals were to evaluate the safety and preliminary effectiveness of the device and refine patient selection criteria.

Methods

Five centers in the United States enrolled 30 patients with juxtarenal aortic aneurysms with ≥50-mm diameter and short proximal necks. Devices were custom-designed for each patient based on measurements from reconstructed computed tomography (CT) data. Follow-up studies included physical examinations, laboratory studies, CT imaging, mesenteric-renal duplex ultrasound imaging, and abdominal flat plate radiographs at hospital discharge, at 1, 6, and 12 months, and yearly thereafter up to 5 years.

Results

During a 1-year period, 30 patients (80% men; mean age, 75 years) with a mean aneurysm size of 61.4 mm were enrolled. In these 30 patients, 77 visceral vessels were accommodated by fenestrations located within the sealing segment of the grafts. The most common design accommodated two renal arteries and the superior mesenteric artery (66.7%). All prostheses were implanted successfully. No visceral arteries were lost. Of the 30 patients treated, 27 were available for 12-month follow-up and 23 were available for 24-month follow-up. No aneurysm-related deaths, aneurysm ruptures, or conversions were observed through 24 months of follow-up. No type I or type III endoleaks were observed. Type II endoleaks were noted in six (26.1%) at 12 months and four (20.0%) at 24 months. No patients had aneurysm growth >5 mm. Aneurysm size decreased in 16 of 23 (69.6%) and was stable in the remaining patients at 24 months. Eight patients experienced a renal event (4 renal artery stenoses, 2 renal artery occlusions, and 2 renal infarcts). Five underwent secondary interventions. No renal failure developed requiring dialysis.

Conclusions

The intermediate-term (24-month) results of the 30 patients in this multicenter study are concordant with previous single-center studies and support the concept that placement of fenestrated endovascular grafts is safe and effective at centers with experience in endovascular repair and renal/mesenteric stent placement.

a Department of Vascular Surgery, Cleveland Clinic Foundation, Cleveland, Ohio

b Oschner Clinic, New Orleans, La

c Division of Vascular Surgery, University of Pittsburgh Medical Center, Pittsburgh, Pa

d Department of Vascular Surgery, Jikei University School of Medicine, Tokyo, Japan and Department of Surgery, Albert Einstein College of Medicine, Manhasset, NY

e Department of Vascular Surgery, University of California, San Francisco, Calif

f MED Institute Inc, West Lafayette, Ind

Corresponding Author InformationReprint requests: Roy K. Greenberg, MD, Director of Endovascular Research, Cleveland Clinic Foundation, Desk S-40, Cleveland, OH 44195

 Competition of interest: Drs Greenberg and Chuter disclose they have intellectual property licensed to Cook Inc and have received travel expenses.

 Additional material for this article may be found online at www.jvascsurg.org.

PII: S0741-5214(09)01187-2

doi:10.1016/j.jvs.2009.05.051


View previous. 9 of 103 View next.