The Powerlink system for endovascular abdominal aortic aneurysm repair: Six-year results
-
The Kaplan-Meier survival curves for the Powerlink and control groups are presented in the figure below. The log-rank test was not statistically significant with P = .5065. The observed survival rates
The Kaplan-Meier survival curves for the Powerlink and control groups are presented in the figure below. The log-rank test was not statistically significant with P = .5065. The observed survival rates and 95% confidence limits of the rates are presented in Table IX.
-
Kaplan-Meier curve of freedom from major adverse events through 72 months in Powerlink and control groups. The log-rank test of the difference in freedom from major adverse event curves between the twKaplan-Meier curve of freedom from major adverse events through 72 months in Powerlink and control groups. The log-rank test of the difference in freedom from major adverse event curves between the two treatment groups was not statistically significant (P =.5251).
-
A, Aneurysm sac diameter reduction over time in Powerlink group. Error bars represent standard deviation. B, Aneurysm sac volume reduction over time in Powerlink group. Error bars represent standard dA, Aneurysm sac diameter reduction over time in Powerlink group. Error bars represent standard deviation. B, Aneurysm sac volume reduction over time in Powerlink group. Error bars represent standard deviation.
Additional material for this article may be found online at www.jvascsurg.org.
Competition of interest: Dr Carpenter owns shares in Endologix, Inc.