Journal of Vascular Surgery
Volume 48, Issue 3 , Pages 546-554.e2, September 2008

Pivotal results of the Medtronic Vascular Talent Thoracic Stent Graft System: The VALOR Trial

Presented at the Annual Meeting of the Society for Vascular Surgery, Baltimore, Md, June 6-9, 2007.

  • Ronald M. Fairman, MD

      Affiliations

    • Hospital of the University of Pennsylvania, Philadelphia, Pa
    • Corresponding Author InformationReprint requests: Ronald M. Fairman, MD, Hospital of the University of Pennsylvania 3400 Spruce St, 4 Silverstein Pavilion, Department of Surgery, Philadelphia, PA 19104
  • ,
  • Frank Criado

      Affiliations

    • Union Memorial Hospital, Baltimore, Md
  • ,
  • Mark Farber, MD

      Affiliations

    • University of North Carolina, Chapel Hill, NC
  • ,
  • Christopher Kwolek, MD

      Affiliations

    • Massachusetts General Hospital, Boston, Mass
  • ,
  • Manish Mehta, MD

      Affiliations

    • Albany Medical Center, Albany, NY
  • ,
  • Rodney White, MD

      Affiliations

    • Harbor–University of California, Los Angeles Medical Center, Torrance, Calif
  • ,
  • Anthony Lee, MD

      Affiliations

    • University of Florida, Gainesville, Fla
  • ,
  • J. Michael Tuchek

      Affiliations

    • Loyola University, Maywood, Ill
  • ,
  • VALOR Investigators

Received 3 December 2007; accepted 29 March 2008. published online 24 June 2008.

Article Outline

Objective

This report summarizes the 30-day and 12-month results of endovascular treatment using the Medtronic Vascular Talent Thoracic Stent Graft System (Medtronic Vascular, Santa Rosa, Calif) for patients with thoracic aortic aneurysms (TAA) who are considered candidates for open surgical repair.

Methods

The study was a prospective, nonrandomized, multicenter, pivotal trial conducted at 38 sites. Enrollment occurred between December 2003 and June 2005. Standard follow-up interval examinations were prescribed at 1 month, 6 months, 1 year, and annually thereafter. These endovascular results were compared with retrospective open surgical data from three centers of excellence.

Results

The Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) trial enrolled 195 patients, and 189 were identified as retrospective open surgical subjects. Compared with the open surgery group, the VALOR test group had similar age and sex distributions, but had a smaller TAA size. Patients received a mean number of 2.7 ± 1.3 stent graft components. The diameters of 25% of the proximal stent graft components implanted were <26 mm or >40 mm. Left subclavian artery revascularization was performed before the initial stent graft procedure in 5.2% of patients. Iliac conduits were used in 21.1% of patients. In 33.5% of patients, the bare spring segment of the most proximally implanted device was in zones 1 or 2 of the aortic arch. In 194 patients (99.5%), vessel access and stent graft deployment were successful at the intended site. The 30-day VALOR results included perioperative mortality, 2.1%; major adverse advents, 41%; incidence of paraplegia, 1.5%; paraparesis, 7.2%; and stroke, 3.6%. The 12-month VALOR results included all-cause mortality, 16.1%; aneurysm-related mortality, 3.1%; conversion to open surgery, 0.5%; target aneurysm rupture, 0.5%; stent graft migration >10 mm, 3.9%; endoleak (12.2%), stent graft patency, 100%; stable or decreasing aneurysm diameter, 91.5%; and loss of stent graft integrity, four patients. No deployment-related events or perforation of the aorta by a graft component occurred. The Talent Thoracic Stent Graft showed statistically superior performance with respect to acute procedural outcomes (P < .001), 30-day major adverse events (41% vs 84.4%, P < .001), perioperative mortality (2% vs 8%, P < .01), and 12-month aneurysm-related mortality (3.1% vs 11.6%, P < .002) vs open surgery.

Conclusions

The pivotal VALOR 12-month trial results demonstrate that the Medtronic Talent Thoracic Stent Graft System is a safe and effective endovascular therapy as an alternative to open surgery in patients with TAA who were considered candidates for open surgical repair.

 

Conventional open repair of a descending thoracic aortic aneurysm (TAA) remains a major invasive surgical procedure with significant inherent risk. These operations require thoracotomy, aortic clamping, partial aortic bypass to support the circulation, and considerable blood loss with associated transfusions. The surgical mortality rate may approach 12% even when performed by experienced surgeons in patients with good cardiac reserve and who are deemed excellent surgical candidates.1 Perioperative morbidity in this referenced cohort included spinal cord ischemia in 14%, respiratory failure in 20%, and renal insufficiency in 13% of patients undergoing open repair. These results have been achieved in centers of excellence using an experienced multidisciplinary team both intraoperatively as well as postoperatively, and may not be widely reproducible.

Physicians have embraced thoracic endograft technology with greater zeal than after United States Food and Drug Administration (FDA) approval of the first endovascular devices to treat abdominal aortic aneurysms. The Medtronic Vascular Talent Thoracic Stent Graft System (Medtronic Vascular, Santa Rosa, Calif) is a minimally invasive endovascular device that offers an alternative treatment for patients with TAA. This report summarizes the pivotal 30-day and 12-month results of the Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) trial, a study designed to evaluate the Talent Thoracic Stent Graft System in patients with TAAs. These endovascular results are compared with retrospective open surgical data from three centers of excellence: The Cleveland Clinic Foundation (Cleveland, Ohio), Massachusetts General Hospital (Boston, Mass), and The Hospital of the University of Pennsylvania (Philadelphia, Pa).

Back to Article Outline

Methods 

Enrollment 

The VALOR trial was a prospective, nonrandomized, multicenter clinical study conducted in the United States to evaluate the safety and efficacy of the Medtronic Vascular Talent Thoracic Stent Graft in the treatment of thoracic aortic diseases. Enrollment occurred from December 2003 to June 2005 at 38 institutions across the United States (Appendix 1, online only). This report focuses on the pivotal test group population, which included patients diagnosed with TAAs. These patients were considered candidates for open surgical repair and were low to moderate risk (0, 1, and 2) per the modified Society for Vascular Surgery and the American Association for Vascular Surgery criteria (Appendix 2, online only).2 The anatomic and medical inclusion and exclusion criteria are presented in Table I.

Table I. Anatomic and medical inclusion and exclusion criteria
Inclusion criteria

Age between 18 and 85

SVS/AAVS criteria 0, 1, or 2

Women with negative pregnancy test 7 days before implant

Fusiform focal TAA ≥5 cm or ≥2 times nonaneurysmal aorta and/or focal saccular TAA or penetrating atherosclerotic ulcer

TAA 20 mm distal to origin of left common carotid artery and 20 mm proximal to the origin of the celiac artery

Proximal and distal neck diameter 18 to 42 mm

Proximal and distal aneurysm neck length >20 mm

TAA confirmed by CTA/MRA with optional three-dimensional reconstruction 3 months before screening

Subject must be able and willing to undergo follow-up imaging and examinations at 1, 6, and 12 months, and annually thereafter

Exclusion criteria

Planned placement of the covered portion of the stent graft in zones 0 or 1

Access vessel precludes safe insertion of the delivery system

Planned aortic conduit

TAA with contained rupture

Connective tissue disease (eg, Marfan syndrome, medial degeneration)

Mycotic aneurysm or is suspected of having systemic infection

Previous stent and/or stent graft or previous surgical repair in the DTA

Treatment of an infrarenal aneurysm at the time of implant

Previous surgical or endovascular treatment of an infrarenal aortic aneurysm

History of bleeding diathesis, coagulopathy, or refuses blood transfusions

Vascular interventional procedure or major surgery 30 days before enrollment

Planned vascular interventional procedure or major surgery ≤30 days of the implant procedure

Cerebrovascular accident ≤3 months

Currently participating in an investigational drug or device clinical trial

Known allergy or intolerance to the device components

Known hypersensitivity or contraindication to anticoagulants or contrast media, which is not amenable to pretreatment

Significant and/or circumferential aortic mural thrombus at proximal or distal attachment sites

Medical condition that may cause noncompliance with the protocol, confound the data interpretation, or a limited life expectancy of <1year

CTA, Computed tomography angiography; DTA, descending thoracic aorta; MRA, magnetic resonance angiography; SVS/AAVS, Society for Vascular Surgery/American Association for Vascular Surgery; TAA, thoracic aortic aneurysm.

Surgical candidates with a fusiform thoracic aortic aneurysm ≥5 cm or ≥2 times the diameter of the nonaneurysmal aorta, as well as focal saccular thoracic aneurysms (penetrating atherosclerotic ulcers), were considered for inclusion. The aneurysm had to be at least 20 mm distal to the left common carotid and 20 mm proximal to the celiac artery, have a proximal and distal nonaneurysmal aortic neck diameter of between 18 and 42 mm, and proximal and distal nonaneurysmal aortic neck lengths of at least 20 mm. A notable exclusion criterion was previous surgical or endovascular treatment of an infrarenal aortic aneurysm.

Device description and deployment 

The Talent Thoracic Stent Graft System consists of a preloaded stent graft and the CoilTrac delivery system (Medtronic). The implanted endoprosthetic portion of the Talent system is composed of a polyester graft fabric sewn to a self-expanding nickel-titanium (chemically polished nitinol) wire frame (Fig 1). Catalog and custom stent graft configurations were made available for this trial (Appendix 3, online only). Stent graft oversizing of 2 to 4 mm relative to the native aortic diameter (measured as adventitia to adventitia) was recommended to provide the necessary outward radial force, maintaining stent graft apposition against the aortic wall. Preoperative diameter and length measurements were obtained from detailed computed tomography angiography (CTA) or magnetic resonance angiography (MRA). The overall design concept is modular, such that additional main sections as well as proximal and distal extensions are introduced separately and mated in vivo as needed to complete the exclusion of the TAA.

  • View full-size image.
  • Fig 1. 

    A, The Talent Thoracic Stent Graft Device consists of a polyester graft fabric sewn to a self-expanding nickel-titanium (chemically polished nitinol) wire frame. B, The modular design concept of the Talent Thoracic Stent Graft Device is illustrated.

The loaded delivery system is inserted in the femoral or iliac artery, tracks through the vasculature, and delivers the stent graft at the target site. Deployment of the proximal stent graft occurs as the outer sheath is withdrawn, initially exposing the proximal bare spring and first covered stent graft. A minimum overlap of 30 mm was required for multiple stent grafts, and the Reliant balloon catheter (Medtronic) could be used at the discretion of the physician to model the stent graft against the aortic wall. Adjunctive surgical techniques, including iliac artery conduits, spinal drains, and left subclavian artery (LSA) revascularization were left to the discretion of the investigator.

The study protocol was approved by the FDA and site institutional review boards. Patients signed an informed consent before participation in the investigational study.

Follow-up protocol 

Standard follow-up evaluations were performed at 1, 6, and 12 months, and annually thereafter. Follow-up visits included a CT scan, chest radiograph, and physical examination. All clinical data were reported by the investigative center on case report forms and monitored by the sponsor. A Clinical Events Committee adjudicated major adverse events (MAEs) for device and procedure relatedness. Medical Metrx Solutions (M2S; West Lebanon, NH) served as the imaging core laboratory and provided critical and comprehensive data evaluation of all imaging studies, ensuring third-party assessment of graft effectiveness.

Endoleaks were defined according to the well-established type I to IV nomenclature.3 In the event the core laboratory could not identify the source, the endoleak was classified as unknown. Migration was defined as >10 mm proximal or distal movement of the stent graft relative to fixed anatomic landmarks, and aneurysm expansion was defined as >5-mm increase in diameter from the 1-month to 12-month follow-up visit. An MAE was defined as death due to the procedure, any death ≤30 days of the procedure, respiratory complications, renal insufficiency or failure, cardiac events, neurologic events, aneurysm rupture, bowel ischemia, major bleeding, or vascular complications. A MAE that was identified as a serious adverse event by the clinical investigator was defined as serious MAE.

Aneurysm-related death was defined as any death ≤30 days from initial implantation or occurring as a consequence of an aneurysm rupture, a conversion to open repair, or any other secondary endovascular procedure relative to the aneurysm that was treated by the Talent Thoracic Stent Graft System as evidenced by CT scan, angiography, or direct observation at surgery or autopsy. Excluded were aneurysms in anatomic areas other than the targeted segment treated by the Talent Thoracic Stent Graft System. Aneurysm-related death after open repair included any death ≤30 days from the surgical procedure or any death caused by reintervention of the targeted aortic segment, or by complications related to the graft or the procedure.

Summary statistics presented for categoric variables are the number in each category and the percentage of known values that this number represents. For continuous variables, the mean and standard deviation are provided; P values were calculated using standard t tests. In some cases the median is provided as well. Kaplan-Meier curves were used to plot freedom from event over time.

Back to Article Outline

Results 

Demographics 

The VALOR trial enrolled 195 patients. Fig 2 details patient accountability of the 195 test group patients at 12 months of follow-p. A total of 189 patients were identified as retrospective open surgical subjects. Subject demographics, baseline history, and aneurysm dimensional characteristics for VALOR test group and the open surgery group are presented in Table II, Table III, Table IV. Compared with the open surgery group, the subjects in the VALOR test group had similar age and gender distributions but had lower TAA size and were less likely to have a previous abdominal aortic aneurysm (AAA) or AAA repair. At the time of enrollment, 51 of 195 patients (26%) in the VALOR test group had aneurysm-related symptoms. The mean aneurysm length was 121.4 ± 72.7 mm as measured by the core laboratory. These data were not available for the open surgery group.

  • View full-size image.
  • Fig 2. 

    Flow chart shows follow-up for participants in the participants Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) trial.

Table II. Subject demographics: VALOR test group vs open surgery
VariableVALOR test groupOpen surgery
Age
Total Population
N195189
Mean ± SD (years)70.2 ± 11.169.6 ± 9.1
Median73.071.0
Min-Max27-8627-85
Male
N11599
Mean ± SD (years)69.3 ± 11.769.9 ± 8.5
Median72.071.0
Min-Max27-8540-84
Female
N8090
Mean ± SD (years)71.6 ± 10.169.3 ± 9.8
Median74.071.0
Min-Max38-8627-85
Sex, % (No.)
Males59.0(115)52.4(99)
Females41.0(80)47.6(90)
Ethnicity, % (No.)
White, non-Hispanic83.1(162)93.7(177)
Black, non-Hispanic12.8(25)5.8(11)
Hispanic (white or black)2.6(5)0.5(1)
Asian/Pacific Islander1.0(2)0(0)
Native American0(0)0(0)
Othera0.5(1)0(0)

SD, standard deviation; VALOR, Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms.

aOne subject had ethnicity specified as “none given.”

Table III. Baseline medical history: VALOR test group vs open surgery
Medical historyVALOR test group % (m/n)Open surgery % (m/n)
Cardiovascular
Angina14.4(28/195)22.8(26/114)
Arrhythmias26.7(52/195)20.3(37/182)
Congestive heart failure8.7(17/195)11.2(21/187)
CABG10.3(20/195)13.3(25/188)
Coronary artery disease40.5(79/195)49.2(91/185)
Hypertension87.2(170/195)88.8(166/187)
Myocardial Infarction13.8(27/195)20.9(39/187)
Peripheral vascular disease16.4(32/195)37.4(70/187)
AAA19.0(37/195)37.0(70/189)
AAA repair2.1(4/195)27.5(52/189)
Gastrointestinal conditions53.8(105/195)NA
Renal insufficiency17.4(34/195)16.0(30/187)
Musculoskeletal conditions53.8(105/195)NA
Neurologic
Cerebral vascular accident9.7(19/195)13.4(25/186)
Paraplegia1.0(2/195)0.5(1/186)
Paraparesis0.5(1/195)NA
Transient ischemic attack7.7(15/195)NA
Pulmonary
COPD36.9(72/195)42.6(80/188)
Tobacco use76.9(150/195)75.9(142/187)
Other abnormal body systems
Hyperlipidemia43.6(85/195)NA
Diabetes15.9(31/195)8.6(16/187)
Bleeding disorders2.6(5/195)NA

AAA, Abdominal aortic aneurysm repair; CABG, coronary artery bypass grafting; COPD, chronic obstructive pulmonary disease; NA, not available; VALOR, Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms.

Table IV. Baseline maximum aneurysm diameters: VALOR test group vs open surgery
Aneurysm diameter, mmSite reported, % (m/n)aCore lab reported, % (m/n)bOpen surgery, % (m/n)
10-170(0/188)0(0/187)0(0/189)
18-290(0/188)0.5(1/187)0(0/189)
30-394.3(8/188)7.5(14/187)0(0/189)
40-4910.6(20/188)20.3(38/187)0.5(1/189)
50-5934.6(65/188)34.8(65/187)13.8(26/189)
60-6933.5(63/188)24.6(46/187)40.7(77/189)
70-7912.2(23/188)10.2(19/187)24.3(46/189)
80-893.2(6/188)2.1(4/187)16.9(32/189)
90-991.1(2/188)0(0/187)0.5(1/189)
100-1090.5(1/188)0(0/187)1.6(3/189)
110-1190(0/188)0(0/187)0.5(1/189)
120+0(0/188)0(0/187)1.1(2/189)

VALOR, Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms.

aDenominator is 188 subjects with site reported data.

bDenominator is 187 subjects with evaluable scans.

Procedure and hospital course 

Vessel access and deployment of the study device at the intended site was successful in 194 (99.5%) of the 195 patients enrolled in the VALOR trial. One patient did not receive a study device because of access failure. Iliac conduits were required for arterial access in 21.1% of the patients. A mean number of 2.7 ± 1.3 stent graft devices (range, 1-7) were implanted per patient. Approximately 25% of the patients had proximal main Talent Thoracic Stent Graft components implanted with diameters <26 mm (3 patients, 1.9%) or >40 mm (49 patients, 23.2%). The highest implantation zone (Fig 3) of the bare spring segment of the most proximally implanted device was zone 1 in 6.7% of patients, zone 2 in 26.8%, zone 3 in 35.6%, and zone 4 in 30.9%.

The decision to revascularize the LSA was left to the implanting physician and was performed before the initial stent graft procedure in 10 of 194 patients (5.2%).4 At the conclusion of the procedure, the 194 patients with a device implanted had patent stent grafts, with integrity maintained and freedom from twisting or kinking.

Clinical utility measures for the VALOR test group and the open surgery group are compared in Table V. The VALOR test group showed superiority in regards to subjects requiring blood transfusion, procedural blood loss, and length of procedure, as well as intensive care unit and overall hospital stay (P < .001).

Table V. Acute procedural data: VALOR test group vs open surgery
VariableVALOR test groupOpen surgeryPa
Subjects requiring blood transfusion, % (m/n)22.7 (44/194)93.7 (164/175)<.001
Blood loss during procedure, mean ± SD mLb371.2±514.43054.9±1702.4<.001
Duration of implant procedure, mean ± SD min154.2±76.0303.3±97.6<.001
Length of stay, mean ± SD
ICU (accessible subjects), hours46.8±114.3185.3±204.7<.001
Overall hospital, days6.4±11.516.7±15.0<.001

SD, Standard deviation; VALOR, Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms.

aFor difference between groups. Percentage of subjects requiring blood transfusion were compared using the Fisher exact test. Other variables were compared using the Wilcoxon test.

bOnly one open surgical site could provide blood loss during procedure data.

Mortality 

Four of 195 VALOR patients (2.1%) died ≤30 days after implantation. Causes of death for these patients included atheroembolic multisystem failure, stroke, periprocedural cardiac arrest, and complications from a myocardial infarction and perforated ulcer. Table VI describes the 30-day mortality rates for the VALOR test group compared with the open surgery group. The VALOR test group experienced a significantly lower rate of early mortality (2% vs 8%, P < .01). All-cause mortality at 12 months is presented in Table VII (16.1% vs 20.6%, P = NS). Freedom from all-cause mortality is presented for both groups in Fig 4. Predictors of all-cause mortality at 12 months in the VALOR patients included prior stroke, with an odds ratio of 4.45 (P = .019), chronic obstructive lung disease, with an odds ratio of 3.72 (P = .008), and aneurysm length, with an odds ratio of 1.008 (P = .017) for each additional millimeter.

Table VI. All-cause mortality at 30-days
Group30-day mortality, % (m/n)a
VALOR test group2.1(4/195)
Open surgery7.9(15/189)

VALOR, Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms.

aP < .01.

Table VII. All-cause mortality at 12 months
Group12 Month Mortality, % (m/n)a
VALOR test group16.1(31/192)
Open surgery20.6(39/189)

VALOR, Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms.

aP = .29.

  • View full-size image.
  • Fig 4. 

    Kaplan-Meier plot of freedom from all-cause mortality for Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) trial participants (solid line) and the open surgery cohort (dashed line).

Aneurysm-related mortality 

Six of 192 patients (3.1%) in the VALOR test group died of an aneurysm-related cause through 12 months of follow-up. Four patients died ≤30 days of the procedure. Two additional late deaths were adjudicated as aneurysm-related. In the open surgery group, 22 of 189 patients (11.6%) died of aneurysm-related causes, and this difference was statistically significant at P < .002. Freedom from aneurysm-related death for both groups is presented in Fig 5.

  • View full-size image.
  • Fig 5. 

    Kaplan-Meier plot of freedom from aneurysm-related mortality for Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) trial participants (solid line) and the open surgery cohort (dashed line).

Conversion to surgery 

One patient (0.5%) was converted to open surgical repair approximately 9 months after implantation for complications related to an apparent infection in the stented segment of the aorta. This patient was alive and fully evaluable at the 12-month postimplantation follow-up.

Major adverse events 

One or more MAEs occurred in 41% (80 of 195) of the VALOR patients ≤30 days after implantation compared with 84.4% (151 of 179) in the open surgery group (P < .001; Table VIII). Most of the individual MAE categories in the endovascular group were lower, but vascular complications were higher in the VALOR patients, at 21% (41 of 195), compared with the open surgery patients, at 12.3% (22 of 179). Freedom from MAEs is presented in Fig 6.

Table VIII. Major adverse events for VALOR test group vs open surgery group at 30 days
CategoryVALOR test group % (m/n)Open surgery % (m/n)
Respiratory complications13.3(26/195)46.9(84/179)
Pneumonia9.2(18/195)22.3(40/179)
Pulmonary embolism0.5(1/195)0.6(1/179)
Pulmonary edema2.1(4/195)24.6(44/179)
Respiratory failure6.2(12/195)26.8(48/179)
Renal complications6.2(12/195)29.1(52/179)
Renal insufficiency1.5(3/195)16.2(29/179)
Renal failure4.6(9/195)19.6(35/179)
Cardiac complications12.3(24/195)44.7(80/179)
Myocardial infarction1.5(3/195)5.6(10/179)
Unstable angina0.5(1/195)0.6(1/179)
New arrhythmia8.7(17/195)41.3(74/179)
Exacerbation of CHF3.1(6/195)5.6(10/179)
Neurologic complications11.8(23/195)20.1(36/179)
New CVA/embolic events3.6(7/195)7.3(13/179)
Paraplegia1.5(3/195)3.4(6/179)
Paraparesis7.2(14/195)12.8(23/179)
Gastrointestinal complications1.0(2/195)0.6(1/179)
Bowel ischemia1.0(2/195)0.6(1/179)
Bleeding complications15.4(30/195)48.0(86/179)
Coagulopathy5.6(11/195)20.1(36/179)
Procedural/Postprocedural14.4(28/195)37.4(67/179)
Vascular complications21.0(41/195)12.3(22/179)
Expanding hematoma at access site1.5(3/195)2.2(4/179)
Pseudo/false aneurysm at access site2.1(4/195)1.1(2/179)
AV fistula0.5(1/195)0(0/179)
Retroperitoneal bleed2.6(5/195)1.1(2/179)
Thrombosis0(0/195)6.1(11/179)
Arterial occlusion2.1(4/195)2.2(4/179)
Vessel rupture/dissection6.2(12/195)1.7(3/179)
Vessel disruption7.7(15/195)0.6(1/179)
Embolism5.1(10/195)1.1(2/179)
Re-op for limb ischemia1.0(2/195)0.6(1/179)
Vascular surgical repair or ultrasound compression required14.4(28/195)3.4(6/179)
Target lesion aneurysm rupture0(0/195)0.6(1/179)
Total major adverse events41.0(80/195)84.4(151/179)

CHF, congestive heart failure; CVA, cerebrovascular accident; VALOR, Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms.

  • View full-size image.
  • Fig 6. 

    Kaplan-Meier plot of freedom from major adverse events at 30 days for Evaluation of the Medtronic Vascular Talent Thoracic Stent Graft System for the Treatment of Thoracic Aortic Aneurysms (VALOR) trial participants (solid line) and the open surgery cohort (dashed line).

Cerebrovascular accidents 

Seven VALOR patients (3.6%) had a periprocedural stroke. Three patients had resolution of stroke-related disability at 12 months, death, or last follow-up. Logistic regression analysis was performed on the occurrence of stroke ≤30 days after the implantation procedure. Patients who had a history of AAA had an odds ratio of 7.1 for the occurrence of stroke (P = .031), and implantation in zone 1 or zone 2 had an odds ratio of 15.2 for the occurrence of stroke (P = .018).

Spinal ischemia 

Postoperative paraplegia occurred ≤30 days in three of 195 VALOR patients (1.5%) and in a fourth patient at 32 days after implantation. All patients had placement of a lumbar drain at the time neurologic deficits were identified. None of these patients experienced recovery at the 1-year follow-up or by the time of death, and none of the patients with paraplegia had a previously treated AAA. Onset of paraparesis occurred ≤30 days in 14 VALOR patients (7.2%). The proportion of patients with unresolved paraparesis within 12 months or last known follow-up fell to 3.1% (6 of 192).

Logistic regression analysis was performed on the incidence of paraplegia or paraparesis within ≤30 days after the implantation procedure. The only covariate that was found to be a significant predictor was the use of a conduit for access, with an odds ratio of 4.13 (P = .020).

Stent graft effectiveness 

The core laboratory identified seven patients with a type I endoleak by the 30-day follow-up visit, as noted in Table IX. Most endoleaks were type II. Sixteen patients had 17 additional endovascular procedures, of which two procedures (1.0%) occurred in the 30-day period before discharge, and 15 procedures (8.1%) occurred at 31 to 365 days. Fourteen procedures were performed to resolve an endoleak. One patient had a procedure to resolve migration and to cover a pseudoaneurysm. One patient was treated for an aneurysmal expansion, and one patient was treated for a second aneurysm.

Table IX. Endoleaks at 1 and 12 months (core laboratory)
Device-related eventAt 1-month visit, % (m/n)At 12-month visit, % (m/n)
Endoleak of any size25.9(45/174)12.2(15/123)
Type I4.0(7/174)4.9(6/123)
Type II15.5(27/174)4.9(6/123)
Type III1.7(3/174)0(0/123)
Type IV0(0/174)0(0/123)
Indeterminate4.6(8/174)2.4(3/123)

The core laboratory noted four stent graft migrations ≤12 months. Two migrations involved the proximal end of the graft moving distally, and two involved the distal end of the graft moving proximally. Only one patient required an additional intervention related to the migration. Aneurysm sac diameter was stable or shrinking in 91.4% of patients. In 11 patients (8.5%), the increase in maximal aneurysm diameter was >5 mm during this interval, and seven of these patients had endoleaks during follow-up. No study patient had loss of stent graft patency or instances of compression or collapse of the endograft ≤12 months.

In two patients the core laboratory confirmed stent fractures ≤12 months. Neither patient had adverse events related to these fractures.

Back to Article Outline

Discussion 

The Talent Thoracic Stent Graft System was first implanted in Australia in January 1996 and received the Conformité Européene (CE) mark in April 1998. The original device has undergone two iterative changes leading up to this pivotal clinical trial, including a delivery system change and chemical polishing of the nitinol stent. Most importantly, the device has not been withdrawn from the commercial market for any reasons related to safety or effectiveness.

Data from worldwide experiences with the Talent Thoracic Stent Graft System have been reported in numerous articles, which describe the use of the device in the whole spectrum of thoracic aortic pathologies. Encouraging results in the elective treatment of thoracic aortic dissection were first reported in 1999,5 followed by positive outcomes in the emergency treatment of aortic perforations due to ruptured TAA, type B aortic dissection, and traumatic injury.6, 7, 8 The most comprehensive long-term experience with the device was described by Fattori et al.9 In the United States, Criado et al10 described their 8-year experience with the Talent Thoracic Stent Graft in 111 TAA patients and 75 type B aortic dissection patients, with an average follow-up of 40 months. The investigators found low mortality and morbidity and favorable midterm survival results.10 After FDA approval of the TAG thoracic endoprosthesis (W. L. Gore & Associates, Flagstaff, Ariz) in March 2005, and publication of the phase II multicenter trial results,11 there has been rapid dissemination of this technology.

The Talent Thoracic Stent Graft offers a wider range of diameter options than is currently available in the commercial United States market. Of the patients implanted with diameters <26 mm or >40 mm, 25% would not have been eligible for endovascular repair using commercially available devices owing to diameter sizing constraints.

In 33.5% of patients, the bare spring segment of the most proximally implanted device was in zones 1 or 2 of the aortic arch. The uncovered proximal stent allows for crossing of the great vessels and proximal fixation in the arch without occluding blood flow. There were no instances of asymmetric opening or asymmetric deployment of the proximal bare spring in this pivotal VALOR trial. In addition, there were no instances of erosion or perforation of the aortic wall by the uncovered proximal nitinol stent. Despite concerns about embolic stroke during endovascular maneuvers in the arch, the incidence of perioperative stroke in this series was remarkably low at 3.6%, and nearly half of these patients had resolution of stroke-related disability at 12 months, death, or last follow-up.

Retrograde type A dissections have been reported with other thoracic endovascular devices as well as in association with open thoracic aortic procedures. In the VALOR trial, retrograde type A dissection developed in three of 195 enrolled patients (1.5%). In one of the patients, the core laboratory determined that the proximal neck length was in fact shorter than 10 mm, although this was not recognized by the site investigators. In another patient the proximal neck was also <20 mm, and the core laboratory recognized evidence of a chronic type B dissection, which was an anatomic exclusion for enrollment in the pivotal arm. In the third patient there was a proximal type I endoleak at the end of the procedure and overzealous ballooning, which inadvertently included the proximal bare spring, was the likely etiology. Further discussion of incidence and prevention of retrograde type A dissection during thoracic aortic endovascular procedures is beyond the scope of this article.

Although the delivery systems were mostly 24F and 25F in size, successful vessel access and deployment occurred in 99.5% of cases, with iliac artery conduits used in 21%. A subset analysis failed to reveal any correlation between French size, conduit use, or vascular complications at 30 days. The use of conduits in this study is comparable to that reported in other contemporary series,12, 13 and experience has dictated that vascular access complications are frequent and may result in death. The need for conduits should be anticipated before arterial injury, particularly in elderly women with small, calcified, stenotic external iliac arteries. Because the longest Talent covered device available for this pivotal trial was 116 mm, 2.7 ± 1.3 devices were placed per patient (range, 1-7 devices). Longer covered endografts (ie, 150 and 200 mm) would have resulted in fewer devices placed per patient. Despite this, the incidence of serious vascular complications in the VALOR pivotal trial (Table IV) compares favorably with the Gore TAG phase II multicenter trial in which longer endografts were introduced and deployed through an indwelling 22F or 24F sheath.

The 30-day paraplegia rate was low, and paraparesis was moderately high in the acute phase. Despite the 30-day paraparesis rate, this event carried a reasonably favorable prognosis, as demonstrated by the unresolved paraparesis rate of 3.1% at 12 months or last known follow-up. Strategies potentially mitigating paraplegia, such as spinal drains, were used at the discretion of the investigator when the perceived risk was significant. An interesting observation is that covariate analysis revealed conduit use was predictive of paraplegia or paraparesis. This is consistent with published reports defining retroperitoneal hemorrhage/hematoma, perioperative hypotension, and injury to the external iliac artery as contributing factors for spinal cord ischemia.14 Future studies will need to identify effective perioperative spinal cord monitoring techniques and interventions, as well as postoperative treatment algorithms.

Although a mean number of 2.7 ± 1.3 stent graft components (range, 1-7 devices) were implanted per patient, no junctional or type III endoleaks were detected at the 12-month follow-up. After deployment, there were no instances of loss of patency or stent graft collapse in this cohort of patients with 12-month follow-up. Continued follow-up of these patients will be necessary to document the long-term efficacy of the device; however, several single-center series using the Talent thoracic device have demonstrated durability.15, 16, 17

The VALOR trial results support the use of the Talent Thoracic Stent Graft System as a safe and effective alternative to open surgical repair in patients with descending TAAs. These elderly patients, despite their significant comorbidities, had low mortality at 30 days and 12 months, as well as aneurysm-related mortality at 12 months, supporting a high rate of successful aneurysm treatment. Specifically, the device showed statistically superior performance with respect to acute procedural outcomes, 30-day MAEs, perioperative mortality, and 12-month aneurysm-related mortality compared with open surgery. These data are particularly meaningful given that the open surgery data were derived from high-volume centers with a reputation for surgical excellence and where the best surgical outcomes would be anticipated.

A review of the recent medical literature allows for comparison between the Talent Thoracic Stent Graft experience vs the Gore TAG device based on Kaplan-Meier estimates.18, 19 Similar rates in 30-day and 12-month all-cause mortality and 12-month aneurysm-related mortality rates have been reported. When serious MAE rates are compared through 12 months by organ system, the VALOR test group and the subjects with a Gore TAG device have essentially the same profile of MAE rates. These comparisons demonstrate that despite fundamental differences in stent graft design, the Talent Thoracic Stent Graft as used in the VALOR test group performed in a substantially similar manner to the Gore TAG device when implanted in a similar group of study subjects.

Back to Article Outline

Conclusions 

The Talent Thoracic Stent Graft System has demonstrated reasonable assurance of safety and effectiveness in treating patients with aneurysms of the descending thoracic aorta who are open surgical candidates with low to moderate surgical risk. The VALOR trial data demonstrate that endovascular therapy with this device results in lower perioperative death, 12-month aneurysm-related death, and lower rates of morbidity than open surgery. Vascular complications associated with thoracic aortic endovascular interventions, including vessel disruption, dissection, rupture, and embolism, are higher than in open surgery and demonstrate the need for further refinement of endovascular thoracic delivery systems. Reduced blood loss and time in intensive care and overall hospital length of stay continue to represent additional advantages of endografting vs open surgical repair. The benefits of the device clearly outweigh the risks when considering the clinically significant results of the VALOR Trial conducted in the intended population.

Back to Article Outline

Author contributions 


Conception and design: RF

Analysis and interpretation: RF, FC, MF, CK, RW, AL, MM, MT

Data collection: RF

Writing the article: RF, FC

Critical revision of the article: RF, FC, MF, CK, RW, AL, MM, MT

Final approval of the article: RF, FC, MF, CK, RW, AL, MM, MT

Statistical analysis: RF

Obtained funding: Not applicable

Overall responsibility: RF

Back to Article Outline

Appendices (online only) 

Appendix 1 (online only). Trial sites and principal investigators
Principal investigatorSiteCity
Ronald Fairman, MDHospital of the University of PennsylvaniaPhiladelphia, Pa
Mark Farber, MDUniversity of North CarolinaChapel Hill, NC
Christopher Kwolek, MDMassachusetts General HospitalBoston, Mass
Michael Dake, MDUniversity of VirginiaCharlottesville, Va
Frank Criado, MDUnion Memorial HospitalBaltimore, Md
David Williams, MDUniversity of MichiganAnn Arbor, Mich
Zvonimir Krajcer, MDSt Luke's Episcopal Hospital - HoustonHouston, Tex
Alan Lumsden, MDBaylor College of MedicineHouston, Tex
Manish Mehta, MDAlbany Medical CenterAlbany, NY
Robert Rhee, MDShadyside Hospital - UPMCPittsburgh, Pa
Anthony Lee, MDUniversity of FloridaGainesville, Fla
Sean Lyden, MDCleveland Clinic Foundation - OhioCleveland, Ohio
Jim Swischuk, MDSt Francis HospitalPeoria, Ill
Paul Bove, MDWilliam Beaumont HospitalRoyal Oak, Mich
Rodney White, MDHarbor UCLATorrance, Calif
Edward Garrett, MDBaptist Memorial HospitalMemphis, Tenn
Michael Tuchek, DOLoyola University Medical CenterMaywood, Ill
Kim Hodgson, MDMemorial Medical CenterSpringfield, Ill
Robert Allen, MDPhysician's Regional Medical CenterNaples, Fla
Phillip Allmendinger, MDHartford HospitalHartford, Conn
Mark Bates, MDCharleston Area Medical CenterCharleston, WV
Daniel Benckart, MDAllegheny General HospitalPittsburg, Pa
Tom Bower, MDMayo Clinic- RochesterRochester, Minn
Mark Eskandari, MDNorthwestern Memorial HospitalChicago, Ill
Neal Cayne, MDNYU Vascular AssociatesNew York, NY
Peter Faries, MDNew York Presbyterian Cornell UniversityNew York, NY
Mark Fillinger, MDDartmouth- Hitchcock Medical CenterLebanon, NH
Marc Glickman, MDSentara Norfolk General, Vascular, and Transplant SpecialistsNorfolk, Va
Matthew Jung, MDSurgical Care AssociatesLouisville, Ky
Barry Katzen, MDBaptist Hospital of MiamiMiami, Fla
Lowell Satler, MDWashington Hospital CenterWashington, DC
Richard McCann, MDDuke University Medical CenterDurham, NC
Takao Ohki, MDMontefiore Medical CenterBronx, NY
Venkatesh Ramaiah, MDArizona Heart InstitutePhoenix, Ariz
Timothy Roush, MDCarolinas Medical CenterCharlotte, NC
Gregorio Sicard, MDWashington University School of MedicineSt Louis, Mo
Cary Stowe, MDFlorida HospitalOrlando, Fla
Christopher Zarins, MDStanford UniversityStanford, Calif
Appendix 2 (online only). Modified Society for Vascular Surgery/American Association for Vascular Surgery medical comorbidity grading system
ComponentSociety for Vascular Surgery risk factor
0 (Absent)1 (Mild)2 (Moderate)3 (Severe)
Age, years<5555-6970-85>85
HypertensionNone (cutoff: DBP usually <90 mm Hg)Controlled (cutoff: point DBP <90 mm Hg) with single drugControlled with 2 or more drugsUncontrolled hypertension
CardiacAsymptomatic, with normal ECGAsymptomatic, but with either remote MI by history (6 months), occult MI by ECG, or fixed deficit on dipyridamole thallium or similar scanAny one of the following:
stable angina

no angina but significant reversible perfusion deficit on Dip-Thal

significant silent ischemia (1% of time) on Holter monitoring

EF 25%-45%

controlled ectopy or asymptomatic arrhythmia

Any one of the following:
unstable angina

symptomatic or poorly controlled ectopy or arrhythmia (chronic/recurrent)

poorly compensated or recurrent CHF

EF <25%

MI ≤6 mon with no intervention (CABG, angioplasty or stenting)

PulmonaryAsymptomatic, normal chest x-ray, PFT ≤20% of predictedAsymptomatic or mild dyspnea on exertion, mild chronic parenchymal x-ray changes, PFT 65%-80% of predictedBetween 1 and 3Vital capacity <1.85 L, FEV1 <1.2 L or <35% of predicted, maximal voluntary ventilation <50% of predicted, Pco2 >45 mm Hg, supplemental O2 medically necessary, or pulmonary hypertension

CABG, coronary artery bypass grafting; DBP, diastolic blood pressure; Dip-Thal, dipyridamole thallium scan; ECG, electrocardiogram; EF, ejection fraction; FEV1, forced expiratory volume in 1 second; MI, myocardial infarction; PFT, pulmonary function tests.

Appendix 3 (online only). Catalog stent graft specifications and configurations
VariablesMain sectionsAdditional distal main sectionsProximal extensionsDistal extensions
Diameters (2-mm increments)
Proximal22-4626-4626-4626-46
Distal22-4622-4426-4626-46
Total covered length of device, mma112-116110-11446-5446-54
Proximal configurationsFreeFlob (Bare Spring)OpenWebFreeFlobOpenWeb
DistalconfigurationsClosedWebClosedWebOpenWebBareSpringb

aThe maximum total length cannot exceed 130 mm.

b“Bare Spring” and “FreeFlo” refer to the configuration in which the terminating spring has no fabric coverage. Bare Spring is the term used for devices having a proximal diameter <24 mm, while FreeFlo is the term used for devices having a proximal diameter ≥24 mm. FreeFlo devices feature a support spring to prevent fabric infolding.

Back to Article Outline

References 

  1. Bavaria JE, Appoo JJ, Makaroun MS, Verter J, Yu ZF, Mitchell RS. Endovascular stent grafting versus open surgical repair of descending thoracic aortic aneurysms in low risk patients: a multicenter comparative trial. J Thorac Cardiovasc Surg. 2007;133:369–377
  2. Chaikoff EL, Fillinger MF, Matsumura JS, Rutherford RB, White GH, Blankensteijn JD, et al. Identifying and grading factors that modify the outcome of endovascular aortic aneurysm repair. J Vasc Surg. 2002;35:1061–1066
  3. White GH, Yu W, May J, Chaufour X, Stephen MS. Endoleak as a complication of endoluminal grafting of abdominal aortic aneurysms: classification, incidence, diagnosis, and management. J Endovasc Surg. 1997;4:152–168
  4. Gorich J, Asquan Y, Seifarth H, Kramer S, Kapfer X, Orend KH, et al. Initial experience with intentional stent-graft coverage of the subclavian artery during endovascular thoracic aortic repairs. J Endovasc Ther. 2002;9:S1139–S1143
  5. Nienaber CA, Fattori R, Lund G, Dieckmann C, Wolf W, von Kodoltisch Y, et al. Nonsurgical reconstruction of thoracic aortic dissection by stent-graft placement. N Engl J Med. 1999;340:1539–1545
  6. Balzer JO, Doss M, Thalhammer A, Fieguth H, Moritz A, Vogl AJ. Urgent thoracic aortal dissection and aneurysm: treatment with stent-graft implantation in an angiographic suite. Eur Radiol. 2003;13:2249–2258
  7. Iannelli G, Piscione F, Di Tommaso L, Monaco M, Chiariello M, Spampinato N. Thoracic aortic emergencies: impact of endovascular surgery. Ann Thorac Surg. 2004;77:591–596
  8. Scheinert D, Kranenberg H, Schmidt A, Gummert JF, Nitzsche S, Scheinert S, et al. Endoluminal stent-graft placement for acute rupture of the descending thoracic aorta. Eur Heart Jl. 2004;25:694–700
  9. Fattori R, Nienaber CA, Rousseau H, Beregi J, Heijmen R, Grabenwoger M, et al. Results of endovascular repair of the thoracic aorta with the Talent Thoracic stent graft: the Talent Thoracic Retrospective Registry. J of Thoracic and Cardiovascular Surg. 2006;132:332–339
  10. Criado F, Abul-Khoudoud O, Domer G, McKendrick C, Zuzga M, Clark NS, et al. Endovascular repair of the thoracic aorta: lessons learned. Ann Thor Surg. 2005;80:857–863
  11. Makaroun MS, Dillavou ED, Kee ST, Sicard G, Chaikof E, Bavaria J, et al. Endovascular treatment of thoracic aortic aneurysms: results of the phase II multicenter trial of the GORE TAG thoracic endoprosthesis. J Vasc Surg. 2005;41:1–9
  12. Greenberg R, Resch T, Nyman U, Lindh M, Brunkwall J, Brunkwall P, et al. Endovascular repair of descending thoracic aortic aneurysms: an early experience with intermediate-term follow-up. J Vasc Surg. 2000;31:147–156
  13. Cambria RP, Brewster DC, Lauterbach SR, Kaufman JL, Geller S, Fan C, et al. Evolving experience with thoracic aortic stent graft repair. J Vasc Surg. 2002;35:1–8
  14. Cheung AT, Pochettino A, McGarvey ML, Appoo JJ, Fairman RM, Carpenter JP, et al. Strategies to manage paraplegia risk after endovascular stent repair of descending thoracic aneurysms. Ann Thor Surg. 2005;80:1280–1289
  15. Jacobs TS, Won J, Gravereau EX, Faries PL, Morrissey N, Teodorescu VJ, et al. Mechanical failure of prosthetic human implants: a 10 year experience with aortic stent graft devices. J Vasc Surg. 2003;37:16–26
  16. Ellozy SH, Carroccio A, Minor M, Jacobs T, Chae K, Cha A, et al. Challenges of endovascular tube graft repair of thoracic aortic aneurysm: midterm follow-up and lessons learned. J Vasc Surg. 2003;38:676–683
  17. Marin ML, Hollier LH, Ellozy SH, Spielvogel D, Mitty H, Griepp R, et al. Endovascular stent graft repair of abdominal and thoracic aortic aneurysms: a ten-year experience with 817 patients. Ann Surg. 2003;238:586–595
  18. Cho JS, Haider S, Makaroun MS. Endovascular therapy of thoracic aneurysms: Gore TAG trial results. Semin Vasc Surg. 2006;19:18–24
  19. Bavaria JE, Appoo JJ, Makaroun MS, Verter J, Yu ZF, Mitchell RS. Endovascular stent grafting versus open surgical repair of descending thoracic aortic aneurysms in low risk patients: a multicenter comparative trial. J Thorac Cardiovasc Surg. 2007;133:369–377

 Supported by Medtronic Vascular.

 Competition of interest: none.

 Additional material for this article may be found online at www.jvascsurg.org.

PII: S0741-5214(08)00511-9

doi:10.1016/j.jvs.2008.03.061